Z-0171-2014 Class II Terminated

Recalled by Handicare Accessibility — Allentown, PA

FDA device recall Z-0171-2014 was initiated by Handicare Accessibility on Nov 29, 2012 and is designated Class II. Reason for recall: Firm became aware that two handsets in the affected batches became inoperable while the user was in the tub. The recall status is terminated (terminated Dec 8, 2014). Affected quantity: 3301 handsets.

Recall details

Product Type
Devices
Report Date
Nov 13, 2013
Initiation Date
Nov 29, 2012
Termination Date
Dec 8, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3301 handsets

Product description

Handset component of the Handicare USA's Minivator Bath Bliss Reclining Bath Lifts (Bath Bliss), Part Number BP300-100/BP300-400RSUS (handset) - Model Numbers BBR0311USAB (blue) and BBR0311USSS (silver) Handset component devices operate the Minivator Bath Bliss Reclining Bath Lift, a non-AC-powered patient lift consisting of a motorized actuator, powered mechanical seating device, temporarily secured in a bath, and used to lift and lower a patient in the vertical or reclining position while bathing.

Reason for recall

Firm became aware that two handsets in the affected batches became inoperable while the user was in the tub.

Distribution pattern

Worldwide distribution: USA (Nationwide) and country of: Canada.

Code information

Handsets shipped with Bath Lifts belonging to the following serial numbers: BH05100001 - BH05129999 and BH06120001 - BH11129999; Handsets shipped as spare parts with the following serial numbers: BS05100001 - BS05129999 and BS06120001 - BS11129999.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0171-2014
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Handicare Accessibility? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works