Z-0171-2018 Class II Terminated

Recalled by Becton Dickinson & Co. — Sparks, MD

Recall Details

Product Type
Devices
Report Date
December 6, 2017
Initiation Date
May 11, 2017
Termination Date
February 6, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
US: 7750 units

Product Description

BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel used to test antibiotic susceptibility. The inducible macrolide-linosamide-steptogramin B (iMLSb) resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. --- Product is packaged 25 panels to 1 box.

Reason for Recall

BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. As the product should not be used when quality control (QC) fails, it is unlikely that there would be any impact on patient results, however, if a customer does not perform QC testing and uses the panel, there is a potential to incorrectly treat a Staphylococcus infection with clindamycin based on the iMLSb test result. This issue can be detected 100% of the time when used per package insert instructions with iMLSb positive strain S. aureus BAA-977.

Distribution Pattern

CT, VA, PA, TX, FL, NC, WI, NM

Code Information

Lot 7031818, Exp. Date 2/28/2018