Z-0171-2018 Class II Terminated
Recalled by Becton Dickinson & Co.
— Sparks, MD
FDA device recall Z-0171-2018 was initiated by Becton Dickinson & Co. on May 11, 2017 and is designated Class II. Reason for recall: BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels… The recall status is terminated (terminated Feb 6, 2018). Affected quantity: US: 7750 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 6, 2017
- Initiation Date
- May 11, 2017
- Termination Date
- Feb 6, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- US: 7750 units
Product description
BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel used to test antibiotic susceptibility. The inducible macrolide-linosamide-steptogramin B (iMLSb) resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. --- Product is packaged 25 panels to 1 box.
Reason for recall
BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. As the product should not be used when quality control (QC) fails, it is unlikely that there would be any impact on patient results, however, if a customer does not perform QC testing and uses the panel, there is a potential to incorrectly treat a Staphylococcus infection with clindamycin based on the iMLSb test result. This issue can be detected 100% of the time when used per package insert instructions with iMLSb positive strain S. aureus BAA-977.
Distribution pattern
CT, VA, PA, TX, FL, NC, WI, NM
Code information
Lot 7031818, Exp. Date 2/28/2018
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0171-2018
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.