Z-0181-2021 Class II Terminated

Recalled by Boston Scientific Corporation — Quincy, MA

FDA device recall Z-0181-2021 was initiated by Boston Scientific Corporation on August 24, 2020 and is designated Class II. Reason for recall: Opticross 35 Peripheral Imaging Catheter is being recalled because the incremental telescoping length markings near the proximal end of the catheter shaft can flake off when the device is manipulated.… The recall status is terminated (terminated April 5, 2024). Affected quantity: 41 devices.

Recall Details

Product Type
Devices
Report Date
October 28, 2020
Initiation Date
August 24, 2020
Termination Date
April 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
41 devices

Product Description

Opticross 35 15 MHz Peripheral Imaging Catheter 8F (2.74 mm) X 105 cm Expiration Date June 8th 2021 - July 22nd 2021 Material Number (UPN) # H7493932800350 UDI # 08714729984542 Lot #25754545; 25852104; 2582106; 25856700 Product Usage: intended for ultrasound examination of peripheral vascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.

Reason for Recall

Opticross 35 Peripheral Imaging Catheter is being recalled because the incremental telescoping length markings near the proximal end of the catheter shaft can flake off when the device is manipulated. No patient harm has been reported to date. The most common potential adverse health consequence would be prolonged procedure to exchange the catheter for another as this potential issue is observable and has occurred outside of the patient's body. The most server health consequence that is reasonably expected to occur if a catheter with this potential issue is used is embolization which would result of the flaking marks advance while flushing the catheter. All recalled devices should be returned to Boston Scientific.

Distribution Pattern

US Nationwide distribution including in the states of (1) MD (2) NY (3) MI (4) CT (5) SD (6) GA.

Code Information

Opticross 35 15 MHz Peripheral Imaging Catheter 8F (2.74 mm) X 105 cm Expiration Date June 8th 2021 - July 22nd 2021 Material Number (UPN) # H7493932800350 UDI # 08714729984542 Lot #25754545; 25852104; 2582106; 25856700