Z-0181-2021 Class II Terminated
Recalled by Boston Scientific Corporation
— Quincy, MA
FDA device recall Z-0181-2021 was initiated by Boston Scientific Corporation on Aug 24, 2020 and is designated Class II. Reason for recall: Opticross 35 Peripheral Imaging Catheter is being recalled because the incremental telescoping length markings near the proximal end of the catheter shaft can flake off when the device is manipulated… The recall status is terminated (terminated Apr 5, 2024). Affected quantity: 41 devices.
Recall details
- Product Type
- Devices
- Report Date
- Oct 28, 2020
- Initiation Date
- Aug 24, 2020
- Termination Date
- Apr 5, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 41 devices
Product description
Opticross 35 15 MHz Peripheral Imaging Catheter 8F (2.74 mm) X 105 cm Expiration Date June 8th 2021 - July 22nd 2021 Material Number (UPN) # H7493932800350 UDI # 08714729984542 Lot #25754545; 25852104; 2582106; 25856700 Product Usage: intended for ultrasound examination of peripheral vascular pathology only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.
Reason for recall
Opticross 35 Peripheral Imaging Catheter is being recalled because the incremental telescoping length markings near the proximal end of the catheter shaft can flake off when the device is manipulated. No patient harm has been reported to date. The most common potential adverse health consequence would be prolonged procedure to exchange the catheter for another as this potential issue is observable and has occurred outside of the patient's body. The most server health consequence that is reasonably expected to occur if a catheter with this potential issue is used is embolization which would result of the flaking marks advance while flushing the catheter. All recalled devices should be returned to Boston Scientific.
Distribution pattern
US Nationwide distribution including in the states of (1) MD (2) NY (3) MI (4) CT (5) SD (6) GA.
Code information
Opticross 35 15 MHz Peripheral Imaging Catheter 8F (2.74 mm) X 105 cm Expiration Date June 8th 2021 - July 22nd 2021 Material Number (UPN) # H7493932800350 UDI # 08714729984542 Lot #25754545; 25852104; 2582106; 25856700
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0181-2021
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Boston Scientific Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Boston Scientific Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Boston Scientific Corporation →
Use it in ChatGPT
Get a free API key
How the connector works