Z-0184-2025 Class II Ongoing
FDA device recall Z-0184-2025 was initiated by Smiths Medical ASD, Inc. on Sep 24, 2024 and is designated Class II. Reason for recall: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar… The recall status is ongoing. Affected quantity: 50 units.
Recall Details
- Product Type
- Devices
- Report Date
- Oct 30, 2024
- Initiation Date
- Sep 24, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 50 units
Product Description
medex NICU Monitoring Kit Set 1/EA, Product Code REF MX20086
Reason for Recall
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Distribution Pattern
Worldwide distribution.
Code Information
UDI/DI 50351688509817 (case), 10351688509819 (pouch), Lot Numbers: 4270117