Z-0185-2019 Class I Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 31, 2018
- Initiation Date
- August 28, 2018
- Termination Date
- December 18, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 30561 units
Product Description
CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use.
Reason for Recall
Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
Distribution Pattern
Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.
Code Information
GTIN: 00380657540037 All lot/serial numbers