Z-0187-2025 Class II Ongoing

Recalled by Smiths Medical ASD, Inc. — Minneapolis, MN

Recall Details

Product Type
Devices
Report Date
October 30, 2024
Initiation Date
September 24, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2232 units

Product Description

medex Arterial Kit 1/EA, Product Code REF MX7827

Reason for Recall

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Distribution Pattern

Worldwide distribution.

Code Information

UDI/DI 50351688505185 (case), 10351688505187 (pouch), Lot Numbers: 4282733, 4319605, 4340472, 4355338, 4390060, 4411161