Z-0189-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- October 30, 2024
- Initiation Date
- September 24, 2024
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1740 units
Product Description
medex TranStar Patient Mount Monitoring Kit 10/EA, Product Code REF MX9501T
Reason for Recall
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Distribution Pattern
Worldwide distribution.
Code Information
UDI/DI 50351688504072 (case), 10351688504074 (pouch), Lot Numbers: 4304896, 4346065, 4355385, 4376424