Z-0191-2018 Class II Terminated

Recalled by CSL Behring GmbH — Marburg, N/A

Recall Details

Product Type
Devices
Report Date
December 6, 2017
Initiation Date
June 5, 2017
Termination Date
April 7, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

SIEMENS N/T Protein Control SL for use on the BN II, BN ProSpec Systems: N/T Protein Control SL/L: REF/Product Code OQIN19, SMN 10446079 (Low); N/T Protein Control SL/M, REF/Product Code OQIO19, SMN 10446085 (Mid); N/T Protein Control SL/H, REF/Product Code OQIP19, SMN 10446089 (High);

Reason for Recall

Siemens Healthcare Diagnostics has confirmed a negative bias of approximately -20% for results obtained with N Antiserum to Human Ceruloplasmin when calibrating Ceruloplasmin with various lots of N Protein Standard SL. N/T Protein Controls SL for Ceruloplasmin are similarly biased. --- The affected standards and controls for the affected lots were incorrectly value assigned, which led to a lower result compared to correctly assigned standards. Note: N Antiserum to Human Ceruloplasmin (REF OUIE) works as intended. --- Siemens investigation indicated that N Protein Standard SL and corresponding N/T Protein Control SL (Levels Low, Mid and High) are inaccurately assigned for Ceruloplasmin.

Distribution Pattern

Worldwide including the countries of United States, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia-Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Finland, France, Germany, Greece, Great Britain, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, japan, Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Mexico, Morocco, Netherlands, New Zealand, Norway, Poland, Portugal, Republic of Korea (S), Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Sweden, South Africa, Switzerland, Syria, Taiwan, Thailand, Trinidad-Tobago, Turkey, Turkmenistan, United Arab Emirates, Uruguay, Vatican, Venezuela,

Code Information

SMN 10446079: Lot (UDI): 084647A (00842768016127084647A20170913), 084647C (00842768016127084647C20170913) exp. date 2017-09-13; 084648(0084276801612708464820180430), 084648C (00842768016127084648C20180430), 084648H (00842768016127084648H20180430), exp. date 2018-04-30. --- SMN 10446085: Lot (UDI): 084749D (00842768016134084749D20170917) exp. date 2017-09-17; 084750 (03 0084276801613408475020180203), 084750B (00842768016134084750B20180203), 084750E (00842768016134084750E20180203) exp. date 2018-02-03. --- SMN 10446089: Lot (UDI) 084847A (00842768016110084847A20170613), 084847B (00842768016110084847B20170613) exp. date 2017-06-13; 084848A (00842768016110084848A20171217), 084848D (00842768016110084848D20171217), 084848E (00842768016110084848E20171217), 084848H (00842768016110084848H20171217), 084848L (00842768016110084848L20171217) exp. date 2017-12-17; 084849 (0084276801611008484920180704), 084849C (00842768016110084849C20180704), 084849D (00842768016110084849D20180704) 084849E (00842768016110084849E20180704) exp. date 2018-07-04.