Z-0192-2018 Class II Terminated

Recalled by HILL-ROM MANUFACTURING, INC. — Saint Paul, MN

Recall Details

Product Type
Devices
Report Date
December 6, 2017
Initiation Date
September 15, 2017
Termination Date
May 29, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
60 devices

Product Description

Monarch Airway Clearance System, Model 1000, Product Numbers: PMACS1NA (has blue tooth capability only) and PMACS1WI (has WIFI capability). The firm name on the label is Hill-Rom Services Pte. Ltd., Singapore. The Monarch Airway Clearance System, Model 1000, is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physicians choice of treatment. It is indicated for patients having difficulty with secretion clearance or the presence of atelectasis caused by mucus plugging.

Reason for Recall

Monarch Electromagnetic Interference (EMI) may disrupt the operation of implanted pacemakers or implantable cardioverter defibrillators as well as other devices that might be subject to an EMI if they come into direct contact with the Monarch device.

Distribution Pattern

USA (nationwide) Distribution to the states of: AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, MA, MI, MN, MO, NE, OH, OK, OR, TN, TX, VA, WA, WI, and WV. Internationally to Canada. There was no military/government distribution.

Code Information

Serial numbers S082VM0092, S122VM0171, S128VM0205, S151VM0242, S082VM0093, S122VM0172, S132VM0210, S151VM0243, S082VM0095, S122VM0175, S132VM0212, S152VM0251, S095VM0119, S122VM0176, S132VM0213, S157VM0257, S095VM0124, S124VM0177, S135VM0216, S157VM0261, S095VM0125, S124VM0178, S136VM0219, S158VM0264, S114VM0152, S124VM0180, S136VM0222, S158VM0266, S119VM0158, S124VM0181, S136VM0223, S158VM0267, S119VM0160, S126VM0187, S138VM0224, S158VM0268, S119VM0161, S126VM0193, S138VM0226, S158VM0271, S119VM0162, S126VM0198, S146VM0227, S158VM0273, S122VM0165, S126VM0199, S146VM0229, S159VM0276, S122VM0166, S128VM0202, S146VM0230, S160VM0279, S122VM0168, S128VM0203, S151VM0238, S167VM0289, S122VM0170, S128VM0204, S151VM0240, and S167VM0291.