Z-0193-2020 Class II Terminated

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

Recall Details

Product Type
Devices
Report Date
October 30, 2019
Initiation Date
July 31, 2019
Termination Date
July 27, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1735600 units

Product Description

BD Vacutainer¿ Push Button Blood Collection Set with Pre-Attached Holder Catalog number 368356 - Product Usage: This product is designed to provide customers an increased level of safety during IV blood collection, convenience and to improve OSHA single-use holder compliance. Figure 1 shows a BD Vacutainer¿ Push Button Blood Collection Set with the needle protector removed prior to needle retraction.

Reason for Recall

Separation of front and rear barrels upon activation of the safety feature that retracts the needle

Distribution Pattern

Worldwide distribution - US Nationwide distribution including the states of CT, FL, MO, TN, OH, TX, GA, NC, SC, IN, IL, PA, CO and countries of Trinidad and Tobago, Hong Kong.

Code Information

Lot / Serial Number. 9079767 9079770 9081829 9086811 9086812 9091607 9091626 9094659 9094661 9098547 9100711 9100712 9115965 9115913