Z-0195-2014 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

FDA device recall Z-0195-2014 was initiated by Terumo Cardiovascular Systems Corporation on Oct 23, 2013 and is designated Class II. Reason for recall: During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of… The recall status is terminated (terminated Jul 18, 2014). Affected quantity: 30 units.

Recall details

Product Type
Devices
Report Date
Nov 20, 2013
Initiation Date
Oct 23, 2013
Termination Date
Jul 18, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
30 units

Product description

Sarns" Soft-Flow¿ Aortic Cannula without Suture Flange, straight tip, wire-reinforced with luer port, aortic cannula, 8.0 mm (24 Fr) OD with 3/8" connector, 14" (36 cm) long. Indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.

Reason for recall

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow¿ Aortic Cannulae and Sarns" Venous Return Cannulae.

Distribution pattern

Worldwide Distribution-USA including DC and the states of MO,TX,OK, LA, MA, CA, WI, and MI and the country of Canada.

Code information

Catalog Number 5841 Lot Numbers 0677302

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0195-2014
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Terumo Cardiovascular Systems Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Terumo Cardiovascular Systems… →

Source: openFDA record for Z-0195-2014. Report a data issue: hello@regdatalab.com.