Z-0197-2015 Class II Terminated
FDA device recall Z-0197-2015 was initiated by Integra LifeSciences Corp. on Oct 24, 2014 and is designated Class II. Reason for recall: A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template… The recall status is terminated (terminated Jul 21, 2015). Affected quantity: 12 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 19, 2014
- Initiation Date
- Oct 24, 2014
- Termination Date
- Jul 21, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 12 units
Product description
Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.
Reason for recall
A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.
Distribution pattern
Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.
Code information
Lot Numbers PM0278 and PM0277