Z-0197-2015 Class II Terminated

Recalled by Integra LifeSciences Corp. — Plainsboro, NJ

FDA device recall Z-0197-2015 was initiated by Integra LifeSciences Corp. on Oct 24, 2014 and is designated Class II. Reason for recall: A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template… The recall status is terminated (terminated Jul 21, 2015). Affected quantity: 12 units.

Recall details

Product Type
Devices
Report Date
Nov 19, 2014
Initiation Date
Oct 24, 2014
Termination Date
Jul 21, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product description

Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder System Cutting Template, 142.5 degrees is an accessory for use with the Integra Titan Reverse Shoulder System. It is used to project the osteotomy angle on the humeral head.

Reason for recall

A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.

Distribution pattern

Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.

Code information

Lot Numbers PM0278 and PM0277

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0197-2015
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0197-2015. Report a data issue: hello@regdatalab.com.