Z-0197-2027 Class II Ongoing

Recalled by Medtronic Navigation, Inc. — Lafayette, CO

FDA device recall Z-0197-2027 was initiated by Medtronic Navigation, Inc. on Aug 31, 2026 and is designated Class II. Reason for recall: Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres. The recall status is ongoing. Affected quantity: 3746 units.

Recall details

Product Type
Devices
Report Date
Oct 28, 2026
Initiation Date
Aug 31, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3746 units

Product description

Medtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly

Reason for recall

Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.

Distribution pattern

US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Canary Islands, Iceland, Northern Ireland, Martinique, China , Taiwan, Hong Kong,

Code information

UDI: 00643169702943/ Lots: 066500526 066501226 066503326 066504726 066506326 066509826; UDI:00643169030121/Lots: 066500526 066506326

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0197-2027
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0197-2027. Report a data issue: hello@regdatalab.com.