Z-0198-2024 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

Recall Details

Product Type
Devices
Report Date
November 8, 2023
Initiation Date
September 6, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5 units

Product Description

RX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: RX4040

Reason for Recall

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

Distribution Pattern

US Nationwide distribution including Puerto Rico

Code Information

GTIN: 05055273207774 Lot/ Batch Number (1) Lot/ Batch Number 7241-0551 7241-0545 7241-0531 7241-0317 7241-0219