Z-0199-2024 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

Recall Details

Product Type
Devices
Report Date
November 8, 2023
Initiation Date
September 6, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product Description

RX Daytona Plus (without ISE)- IVD Clinical Chemistry Analyser (1) RX4041, (2) RX4041R (refurbished)

Reason for Recall

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

Distribution Pattern

US Nationwide distribution including Puerto Rico

Code Information

GTIN: 05055273207781 Lot/Batch: (1) 7241-0616 7241-0582 7241-0441 7241-0425 7241-0384 RX4041R: Lot Number: (2) 7241-0149