Z-0200-2025 Class II Ongoing

Recalled by Smiths Medical ASD, Inc. — Minneapolis, MN

FDA device recall Z-0200-2025 was initiated by Smiths Medical ASD, Inc. on September 24, 2024 and is designated Class II. Reason for recall: Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar … The recall status is ongoing. Affected quantity: 3280 units.

Recall Details

Product Type
Devices
Report Date
October 30, 2024
Initiation Date
September 24, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3280 units

Product Description

medex TranStar KIDS KIT 7INCH CLOSED BLOOD SAMPLE KIT with 30cc/hr Flush Device 10/EA, Product Code REF MX9547

Reason for Recall

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Distribution Pattern

Worldwide distribution.

Code Information

UDI/DI 50351688503013 (case), 10351688503015 (pouch), Lot Numbers: 4223699, 4248326, 4297927, 4323137, 4340498, 4346068, 4350022