Z-0202-2024 Class II Ongoing

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

Recall Details

Product Type
Devices
Report Date
November 8, 2023
Initiation Date
September 6, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 units

Product Description

RX Monaco-IVD Clinical Chemistry Analyser Catalog Number: RX5000

Reason for Recall

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

Distribution Pattern

US Nationwide distribution including Puerto Rico

Code Information

GTIN: 05055273207750 Lot/ Batch Number 230T240CS0126S 210T240CS0258S S180T240CS0486MA S180T240CS0487MA S180T240CS0483MA S180T240CS0218MA S170T240CS0665MA S160T240CS0111MA S160T240CS0110MA S150T240CS0005MA S140T240CS0185MA S140T240CS0017MA S140T240CS0015MA