Z-0207-2027 Class II Ongoing
FDA device recall Z-0207-2027 was initiated by TECHNO-PATH MANUFACTURING LTD. on Aug 14, 2026 and is designated Class II. Reason for recall: A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the… The recall status is ongoing. Affected quantity: 832 units.
Recall details
- Product Type
- Devices
- Report Date
- Oct 28, 2026
- Initiation Date
- Aug 14, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 832 units
Product description
Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
Reason for recall
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Distribution pattern
US Nationwide distribution in the state of Massachusetts.
Code information
UDI:(01)05391523442824/-Lot # 65705263