Z-0207-2027 Class II Ongoing

Recalled by TECHNO-PATH MANUFACTURING LTD. — Ballina Tipperary

FDA device recall Z-0207-2027 was initiated by TECHNO-PATH MANUFACTURING LTD. on Aug 14, 2026 and is designated Class II. Reason for recall: A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the… The recall status is ongoing. Affected quantity: 832 units.

Recall details

Product Type
Devices
Report Date
Oct 28, 2026
Initiation Date
Aug 14, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
832 units

Product description

Techno-Path Multichem AE Unassayed Level 3, REF:AE623X

Reason for recall

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

Distribution pattern

US Nationwide distribution in the state of Massachusetts.

Code information

UDI:(01)05391523442824/-Lot # 65705263

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0207-2027
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0207-2027. Report a data issue: hello@regdatalab.com.