Z-0212-2024 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

FDA device recall Z-0212-2024 was initiated by Baxter Healthcare Corporation on Oct 4, 2023 and is designated Class II. Reason for recall: Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport… The recall status is ongoing. Affected quantity: 499 beds.

Recall details

Product Type
Devices
Report Date
Nov 8, 2023
Initiation Date
Oct 4, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
499 beds

Product description

HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052, P7501A000053, P7501A000054, P7501A000055, P7501A001851, P7501A001972, P7501A001973, and P7501A001974

Reason for recall

Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild static electric shock.

Distribution pattern

US Nationwide Distribution in the states of: CA, FL, IN, MA, NY, OH, TX, VA

Code information

UDI/DI 00887761966925, all serial numbers

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0212-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0212-2024. Report a data issue: hello@regdatalab.com.