Z-0220-2021 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 28, 2020
- Initiation Date
- July 6, 2020
- Termination Date
- May 11, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- Total of all products (Listed #1 thru 101) = 304735 units
Product Description
Appx 0.82 ml, 3 Gang 4-Way NanoClave¿ Stopcock Manifold w/Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709053189(17)241201(30)50(10)4537253;(01)10887709053189(17)250101(30)50(10)4557545. The ICU Medical Clave Stopcock is a device used to administer or withdraw fluids that flow from a container to a patient's vascular system through a needle or catheter inserted into a vein or artery. This device is a four-way stopcock with an integral CLAVE side-port for closed, needleless access to the patient tubing or other fluid container.
Reason for Recall
Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Distribution Pattern
Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Code Information
Lot Numbers:4537253, 4557545.