Z-0223-2018 Class II Terminated
FDA device recall Z-0223-2018 was initiated by Terumo Medical Corp on Jun 7, 2017 and is designated Class II. Reason for recall: An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the affected lot population, 0.01% (1 in 10, 000 units)… The recall status is terminated (terminated Apr 25, 2019).
Recall details
- Product Type
- Devices
- Report Date
- Dec 20, 2017
- Initiation Date
- Jun 7, 2017
- Termination Date
- Apr 25, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
Glidesheath Slender Hydrophilic Coated Introducer Sheath, 5FR - 10CM (0.021), REF/Product Code RM*RS5F10PA, STERILE, Rx ONLY
Reason for recall
An incomplete seal at the chevron end of the primary packaging of the GLIDESHEATH SLENDER Introducer Sheath was detected in a small quantity of the affected lot population, 0.01% (1 in 10, 000 units). This breach of the sterile barrier may compromise the sterility of the sheath.
Distribution pattern
Nationwide and Canada
Code information
Lot s 161019, 161115, 161212, 170112