Z-0230-2016 Class I Terminated

Recalled by Event Medical LTD — Lake Forest, CA

FDA device recall Z-0230-2016 was initiated by Event Medical LTD on Oct 13, 2015 and is designated Class I. Reason for recall: A component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming. The recall status is terminated (terminated Jun 22, 2017). Affected quantity: 6441 units total (251 units in US).

Recall details

Product Type
Devices
Report Date
Nov 25, 2015
Initiation Date
Oct 13, 2015
Termination Date
Jun 22, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6441 units total (251 units in US)

Product description

Inspiration Ventilator System, i Series and LS 12.1 LCD. The Inspiration 7i / 5i Ventilator System is to be used by healthcare professionals in hospitals or healthcare facilities and intra-hospital transport.

Reason for recall

A component on the systems power board can fail in a manner which can cause the ventilator to shut down without alarming.

Distribution pattern

Worldwide Distribution - US (Nationwide) and Internationally to: Algeria Argentina Australia Austria Bahrain Bangladesh Barbados Belarus Belgium Bolivia Canada Chile China Colombia Czech Rep Ecuador Egypt England Estonia Finland Germany Ghana Gibraltar Greece Guatemala Holland Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kenya Korea Kuwait Laos Latin America Lebanon Libya Lithuania Malaysia Mexico Morocco Nepal Netherlands New Zealand Norway Pakistan Palestine Panama Peru Philippines Poland Portugal Russia Saudi Arabia Serbia Singapore South Africa Spain Sri Lanka Sudan Switzerland Syria Taiwan Thailand Turkey UAE Ukraine USA Venezuela Vietnam Yemen Zimbabwe

Code information

All models manufactured prior to 2015 are affected

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0230-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Event Medical LTD? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0230-2016. Report a data issue: hello@regdatalab.com.