Z-0231-2025 Class II Ongoing
FDA device recall Z-0231-2025 was initiated by ImaCor Inc. on Sep 16, 2024 and is designated Class II. Reason for recall: Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained… The recall status is ongoing. Affected quantity: 73 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 6, 2024
- Initiation Date
- Sep 16, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 73 units
Product description
ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number: CLT-010, CLT-010-1
Reason for recall
Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.
Distribution pattern
FL ,GA, NJ, TN
Code information
UDI - Primary - 00861589000108 Lot numbers: 20250509 and 20250823
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0231-2025
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
Need the full FDA record for ImaCor Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.