Z-0232-2018 Class II Terminated

Recalled by Fujifilm Medical Systems U.S.A., Inc. — Stamford, CT

Recall Details

Product Type
Devices
Report Date
December 20, 2017
Initiation Date
June 15, 2017
Termination Date
May 9, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product Description

FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR Visionary Suite is intended to generate digital or conventional radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts of human anatomies in all routine radiography examinations.

Reason for Recall

A potential issue in the X-ray tube ceiling unit CH-200 which constitutes the Digital Radiography System FDR Visionary Suite. The X-ray tube assembly in these devices is mounted to a tube mounting flange, which is part of the tube holding shaft. where cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.

Distribution Pattern

US Distribution.

Code Information

UDI: (01)04540217052226(21)MP95A9F6B001; (01)04540217052226(21)MP95A9F6A001; (01)04540217052226(21)MP95A8F5A001. Serial Numbers: FDR Visionary Suite System S/N MP95A9F6B001 MP95A9F6A001 MP95A8F5A001