Z-0234-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- October 31, 2018
- Initiation Date
- September 14, 2018
- Termination Date
- February 12, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 9,364 cases (46,820 eaches)
Product Description
Disposable Accessory Kit, 4-Arm (Box of 5) individually sealed in poly-Tyvek pouches / Rx Sterile / to used exclusively on the Intuitive Surgical Da Vinci Si Robotic system / Distributed by Intuitive Surgical Product Usage: Microtek Medical Equipment Drapes are intended to cover/wrap various surgical and/or nonsurgical instruments/equipment [e.g. robotic arms, microscopes, tables, x ray systems, light handles, etc]. They function as a physical barrier to prevent cross contamination between the instrument / equipment and medical staff and/or to allow the instrument/equipment to enter a hygienic area [e.g. sterile surgical field] in various clinical settings. They are typically made of flexible plastic and shaped to fit the type of instrument / equipment. They are a single use device [not to be reprocessed]
Reason for Recall
The pouches of certain lots of product may have wrinkles along the pouch seal that could result in a channel within the seal. Should there be a channel in the pouch seal it may result in a breach in the sterility of the product.
Distribution Pattern
Worldwide - US Nationwide Distribution in the states of AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, AND WY In the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Iceland, Ireland, Italy, Martinique, Netherlands, Norway, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, and United Kingdom
Code Information
Item 420291-03 Lot # D173147, D173177, D173517, D173437, D173487, D172937, D173407, D173417, D173317, D180117, D180177, D173477, D180087, D173187, D173137, D172987, D173217, D173067, D173497, D173447, D173427, D173287, D180167, D180027, D173577, D173077, D180097, D180157, D173627, D173607, D173467, D173567, D180197, D173257, D173637, D173457, D180267, D180257, D172997, D173547, D180317, D180237, D173227, D173537, D173507, D180107, D173617, D173267, D180247, D173197, D180057, D180187, D180347, D180337, D180377, D180037, D180307, D180437, D173017,and D180507.