Z-0234-2022 Class I Terminated

Recalled by Arrow International Inc — Reading, PA

Recall Details

Product Type
Devices
Report Date
November 24, 2021
Initiation Date
September 20, 2021
Termination Date
October 5, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1918 units

Product Description

Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-WC The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venousfistulae and synthetic dialysis grafts

Reason for Recall

7FR Percutaneous Thrombolytic Device (PTD) Inner lumen (orange) detached from the device basket result in embolization and retrieval with an additional intervention and consequent delay of therapy.

Distribution Pattern

Worldwide distribution: US (Nationwide) and OUS (Foreign) to countries of: BE CA CL CZ DE ES FR IL IT KR PH SA TR TW UK

Code Information

Lot Number: 13F19J0656 13F19K0368 13F19K0342 13F20C0596 13F20F0081 13F20F0229 13F20F0509 13F20F0578 13F20G0177 13F20G0566 13F20H0531 13F20J0379 13F20L0514 13F21A0354 13F21C0081 13F21C0749 13F21D0870 13F21E0415 13F21F1188