Z-0236-2019 Class II Terminated

Recalled by GE Medical Systems, LLC — Waukesha, WI

FDA device recall Z-0236-2019 was initiated by GE Medical Systems, LLC on Oct 3, 2018 and is designated Class II. Reason for recall: A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used. The recall status is terminated (terminated May 27, 2022). Affected quantity: 4 units.

Recall Details

Product Type
Devices
Report Date
Oct 31, 2018
Initiation Date
Oct 3, 2018
Termination Date
May 27, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4 units

Product Description

SIGNA Architect system Product Usage: The Signa Architect system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Reason for Recall

A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.

Distribution Pattern

Worldwide - US Nationwide Distribution in the state of Utah. International distribution to France, Republic of Korea, and Sweden.

Code Information

Model number: 5388807-100, Serial number (System ID number ): U30RF0251B (801588MR3), U30RF0258B (M4040093), U30RF0275B (850270504), U30RF0268B (A020MR05)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0236-2019
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.