Z-2314-2026 Class II Ongoing

Recalled by GE Medical Systems, LLC — Waukesha, WI

FDA device recall Z-2314-2026 was initiated by GE Medical Systems, LLC on Apr 20, 2026 and is designated Class II. Reason for recall: GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service… The recall status is ongoing. Affected quantity: 9 units.

Recall details

Product Type
Devices
Report Date
Jun 17, 2026
Initiation Date
Apr 20, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
9 units

Product description

Allia IGS 3 Pulse angiographic X-ray system

Reason for recall

GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.

Distribution pattern

Worldwide - US Nationwide distribution.

Code information

UDI-DI: 00195278719577; Serial Numbers: M3-24-053, M3-24-124, M3-25-029, M3-25-051, M3-25-053, M3-25-091, M3-25-098, M3-25-124, M3-26-010, M3-26-030

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2314-2026
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2314-2026. Report a data issue: hello@regdatalab.com.