Z-2318-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 17, 2026
- Initiation Date
- April 20, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 units
Product Description
Allia Moveo angiographic X-ray system
Reason for Recall
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Distribution Pattern
Worldwide - US Nationwide distribution.
Code Information
UDI-DI: 00198953052488; Serial Numbers: C4-25-001, C4-25-002, C4-26-001, C4-26-002, C4-26-004, C4-26-005