Z-0238-2015 Class I Terminated

Recalled by Customed, Inc — Fajardo, PR

FDA device recall Z-0238-2015 was initiated by Customed, Inc on Oct 8, 2014 and is designated Class I. Reason for recall: Customed products are being recalled because of compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes and uncontrolled and… The recall status is terminated (terminated Aug 14, 2017). Affected quantity: 76800.

Recall details

Product Type
Devices
Report Date
Dec 10, 2014
Initiation Date
Oct 8, 2014
Termination Date
Aug 14, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
76800

Product description

Irrigation Tray With Bulb Syringe, Catalog number 900-020. Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

Reason for recall

Customed products are being recalled because of compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.

Distribution pattern

US Distribution to Florida, New York and Puerto Rico..

Code information

Lot numbers: 111112971 111123273 112010006 112020294 112030797 112041052 112041386 112041450 112051644 112062300 112083043 112093635 113025799 113025979 113026274 113036446 113046993 113047330 113057829 113068135 113078452 113078659 113099261 131110143 140111104 140211714 140312032 140412568 140513034 140613550 140814606 140815025

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0238-2015
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Customed, Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Customed, Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Customed, Inc →

Source: openFDA record for Z-0238-2015. Report a data issue: hello@regdatalab.com.