Z-0244-2022 Class II Terminated

Recalled by Laboratoires Bodycad, Inc. — Quebec, N/A

FDA device recall Z-0244-2022 was initiated by Laboratoires Bodycad, Inc. on Jul 2, 2021 and is designated Class II. Reason for recall: There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use. The recall status is terminated (terminated Jan 4, 2023). Affected quantity: 10 in the U.S.

Recall details

Product Type
Devices
Report Date
Nov 24, 2021
Initiation Date
Jul 2, 2021
Termination Date
Jan 4, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10 in the U.S.

Product description

Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.

Reason for recall

There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.

Distribution pattern

Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.

Code information

Torque Limiting Adaptor - Lot number SB000430. Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0244-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Laboratoires Bodycad, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Laboratoires Bodycad, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Laboratoires Bodycad, Inc. →

Source: openFDA record for Z-0244-2022. Report a data issue: hello@regdatalab.com.