Z-0244-2022 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- November 24, 2021
- Initiation Date
- July 2, 2021
- Termination Date
- January 4, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10 in the U.S.
Product Description
Torque Limiting Adaptor, AO Connect, 4 N-m, part #027-203, reusable, a component of the Bodycad Fine Osteotomy Implant and Disposable Instrumentation Tray, part #025-004.
Reason for Recall
There is a risk of breakage of the torque limiting adaptor if used counterclockwise and it must not be used with a power tool due to no test data validating this use.
Distribution Pattern
Distribution was made to CO, MI, and NY. Foreign distribution was made to Canada and the United Kingdom. There was no military/government distribution.
Code Information
Torque Limiting Adaptor - Lot number SB000430. Fine osteotomy tray - Lot numbers SB000251, SB000252, SB000253, SB000254, SB000259, SB000261, SB000263, SB000264, SB000272, and SB000273.