Z-0245-2020 Class II Terminated

Recalled by Alto Development Corp — Wall Township, NJ

FDA device recall Z-0245-2020 was initiated by Alto Development Corp on Sep 10, 2019 and is designated Class II. Reason for recall: Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which occludes the through-channel for cable tensioning, rendering the tensioner inoperable. The recall status is terminated (terminated May 12, 2020). Affected quantity: 209 total.

Recall details

Product Type
Devices
Report Date
Nov 6, 2019
Initiation Date
Sep 10, 2019
Termination Date
May 12, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
209 total

Product description

A&E Medical Thorecon Plating System Kits Model No. 94-1500-04-S. For fracture fixation or reconstruction of the sternum.

Reason for recall

Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which occludes the through-channel for cable tensioning, rendering the tensioner inoperable.

Distribution pattern

Distributed to AL, CA, FL, GA, IL, LA, MO, OK, MS, NM, OK, SC, TX, UT, VT, WA, and WI.

Code information