Z-0245-2022 Class II Ongoing

Recalled by Medicalplastic S.R.L. — Milan, N/A

FDA device recall Z-0245-2022 was initiated by Medicalplastic S.R.L. on Jun 17, 2021 and is designated Class II. Reason for recall: Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been… The recall status is ongoing. Affected quantity: 581 units.

Recall details

Product Type
Devices
Report Date
Nov 24, 2021
Initiation Date
Jun 17, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
581 units

Product description

Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434

Reason for recall

Medicalplastic s.r.l. is conducting a voluntary field action to remove a group of Ilizarov Wire Tensioners due to a manufacturing error. An internal washer of the Ilizarov Wire Tensioner has been made in a wrong plastic material that could wear and break during the sterilization process in autoclave. Consequently this could cause malfunctioning of the tensioner.

Distribution pattern

Memphis, TN

Code information

Lot # 20LP02813, Serial #D147 to D149; Lot # 20LP02812, Serial # D115 to D119 ; Lot # 20LP02816, Serial # D099 to D114 ; Lot # 20HP01926, Serial # D059 to D065 ; Lot # 20LP02819, Serial # D175 to D198 ; Lot # 20LP02818, Serial # D129 to D146 ; Lot # 20LP02815, Serial # D120 to D128 ; Lot # 20MP03078, Serial # D374 ; Lot # 20HP01927, Serial # D278 to D282 ; Lot # 20MP03037, Serial # D294 to D296 ; Lot # 20LP02827, Serial # D199 to D258 ; Lot # 20 MP03036, Serial # D288 to D293 ; Lot # 20MP03077, Serial # D370 to D373 ; Lot # 20MP03035, Serial # D259 to D277 ; Lot # 20MP03041, Serial # D283 to D287 ; Lot # 20MP03049, Serial # D343 to D369 ; Lot # 20MP03048, Serial # D297 to D342 ; Lot # 20MP03123, Serial # D375 to D459 ; Lot # 20HP01928, Serial # D460 to D474 ; Lot # 20MP03126, Serial # D482 to D491 ; Lot # 20MP03125, Serial # D492 to D522 ; Lot # 20MP03124, Serial # D475 to D481 ; Lot # 21BP00265, Serial # D523 to D574 ; Lot # 21BP00266, Serial # D626 to D672 ; Lot # 21BP00271, Serial # D588 to D625 ; Lot # 21BP00272, Serial # D575 to D587 ;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0245-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medicalplastic S.R.L.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medicalplastic S.R.L.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medicalplastic S.R.L. →

Source: openFDA record for Z-0245-2022. Report a data issue: hello@regdatalab.com.