Z-0247-2016 Class II Terminated

Recalled by Maquet Cardiovascular Us Sales, Llc — Wayne, NJ

FDA device recall Z-0247-2016 was initiated by Maquet Cardiovascular Us Sales, Llc on Aug 17, 2015 and is designated Class II. Reason for recall: Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine… The recall status is terminated (terminated Oct 19, 2016). Affected quantity: United States (3 sets); Rest of World (47 sets).

Recall details

Product Type
Devices
Report Date
Nov 18, 2015
Initiation Date
Aug 17, 2015
Termination Date
Oct 19, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
United States (3 sets); Rest of World (47 sets)

Product description

Maquet Getinge Group, Adjustable Positioning Elements (1007.28A0) Manufacturer: Maquet GmbH, Germany. Patient positioners. The MAQUET Adjustable Positioning Elements are designed for the placement and positioning of patients in the prone position immediately before, during and after surgical interventions as well as for examination and treatment. With a patient in the prone position, the adjustable positioning elements stabilize the upper body while simultaneously relieving the abdominal region in the vertical position with regard to the spine and the large blood vessels there. This positioning method is thus suitable for interventions on the spine with dorsal access to prevent positioning-specific compression with resulting bleeding tenancies. They are made up of the following components: (2) chest and (2) hip elements with padding which are secured in a fixed arrangement to the frames and can be adjusted in all directions independently of each other.

Reason for recall

Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elements may have a negative effect on patients- venous backflow under certain conditions and may have been a contributing factor to thromboembolic events.

Distribution pattern

Worldwide Distribution - US including AL, CA, MO and Internationally to Australia, Brazil, Canada, China, Croatia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Russia, Singapore, South Africa and United Arab Emirates.

Code information

Part Nos 1007.28A0 (Adjustable Positioning Elements and 1007.24F0 (Universal Frame with Adjustable Positioning Elements)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0247-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0247-2016. Report a data issue: hello@regdatalab.com.