Z-0247-2016 Class II Terminated
Recalled by Maquet Cardiovascular Us Sales, Llc
— Wayne, NJ
FDA device recall Z-0247-2016 was initiated by Maquet Cardiovascular Us Sales, Llc on Aug 17, 2015 and is designated Class II. Reason for recall: Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine… The recall status is terminated (terminated Oct 19, 2016). Affected quantity: United States (3 sets); Rest of World (47 sets).
Recall details
- Product Type
- Devices
- Report Date
- Nov 18, 2015
- Initiation Date
- Aug 17, 2015
- Termination Date
- Oct 19, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- United States (3 sets); Rest of World (47 sets)
Product description
Maquet Getinge Group, Adjustable Positioning Elements (1007.28A0) Manufacturer: Maquet GmbH, Germany. Patient positioners. The MAQUET Adjustable Positioning Elements are designed for the placement and positioning of patients in the prone position immediately before, during and after surgical interventions as well as for examination and treatment. With a patient in the prone position, the adjustable positioning elements stabilize the upper body while simultaneously relieving the abdominal region in the vertical position with regard to the spine and the large blood vessels there. This positioning method is thus suitable for interventions on the spine with dorsal access to prevent positioning-specific compression with resulting bleeding tenancies. They are made up of the following components: (2) chest and (2) hip elements with padding which are secured in a fixed arrangement to the frames and can be adjusted in all directions independently of each other.
Reason for recall
Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elements may have a negative effect on patients- venous backflow under certain conditions and may have been a contributing factor to thromboembolic events.
Distribution pattern
Worldwide Distribution - US including AL, CA, MO and Internationally to Australia, Brazil, Canada, China, Croatia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Russia, Singapore, South Africa and United Arab Emirates.
Code information
Part Nos 1007.28A0 (Adjustable Positioning Elements and 1007.24F0 (Universal Frame with Adjustable Positioning Elements)
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0247-2016
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.