Z-0248-2021 Class II Terminated

Recalled by ICU Medical, Inc. — San Clemente, CA

Recall Details

Product Type
Devices
Report Date
October 28, 2020
Initiation Date
July 6, 2020
Termination Date
May 11, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Total of all products (Listed #1 thru 101) = 304735 units

Product Description

6.5 IN (17 cm) APPX 0.78 ml, Ext Set w/Clave Clear, Clave 4-Way Stopcock, Spiros Red Cap;1 unit per pouch, 50 pouches per case. UDI:(01)10887709098878(17)241201(30)50(10)4493149. The ChemoClave is a needle-free Closed System Transfer Device (CSTD) that mechanically prohibits the transfer of environmental contaminants, including bacterial and airborne contaminants into the system, and the escape of drug or vapor concentrations outside the system during drug preparation and administration, thereby minimizing exposure of individuals, healthcare personnel and the environment to hazardous drugs.

Reason for Recall

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Distribution Pattern

Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.

Code Information

Lot Number:4493149.