Z-0253-2013 Class III Terminated
FDA device recall Z-0253-2013 was initiated by Siemens Healthcare Diagnostics, Inc. on Aug 3, 2012 and is designated Class III. Reason for recall: Cartridges are leaking. The leaking is observed when opening the wrapper that holds the reagent cartridge. The recall status is terminated (terminated Jan 15, 2014). Affected quantity: 583 cartons.
Recall details
- Product Type
- Devices
- Report Date
- Nov 14, 2012
- Initiation Date
- Aug 3, 2012
- Termination Date
- Jan 15, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 583 cartons
Product description
Siemens Dimension Vista(R) Homocysteine Flex(R) Reagent Cartridge This is an in vitro diagnostic test for the quantitative measurement of total homocysteine (HCYS) in human serum, heparinized plasma, and EDTA plasma on the Dimension Vista(R) System. Measurements of homocysteine aid in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.
Reason for recall
Cartridges are leaking. The leaking is observed when opening the wrapper that holds the reagent cartridge.
Distribution pattern
USA (nationwide) including the states of AL, CA, CO, CT, DC, DE, FL, IA, IL, IN, LA, MA, MD, MI, MO, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA and WI.
Code information
Lot numbers 12062MA, exp 12/01/2012; and 12066MA, exp 12/05/2012