Z-0257-2024 Class II Ongoing

Recalled by Stryker Corporation — Portage, MI

Recall Details

Product Type
Devices
Report Date
November 15, 2023
Initiation Date
September 11, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,321 total units

Product Description

Disposable Spetzler-Malis Bipolar forceps, 23CM SLM DISP BAYONET 1.5MM, Catalog Number 6780-230-015

Reason for Recall

An error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.

Distribution Pattern

US Nationwide. Australia, Canada, Japan, Korea, Singapore.

Code Information

GTIN: 7613327300239; Lot Numbers: 200423023, 200525015, 200806014, 201015018