Z-0257-2027 Class II Ongoing

Recalled by Boston Scientific Cork, Ltd. — Cork

FDA device recall Z-0257-2027 was initiated by Boston Scientific Cork, Ltd. on Sep 16, 2026 and is designated Class II. Reason for recall: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves… The recall status is ongoing. Affected quantity: 1135.

Recall details

Product Type
Devices
Report Date
Oct 28, 2026
Initiation Date
Sep 16, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1135

Product description

Direxion Torqueable Microcatheter, NGMC/FATHOM/027/BERN/1RO/130; Material Number: M001195740

Reason for recall

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

Distribution pattern

Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom.

Code information

Material Number: M001195740; UDI-DI: 08714729839712; Batch Numbers: 38092416, 38112643, 38149830, 38181974, 38224588, 38224589, 38259488, 38273747, 38273748, 38290697, 38330742, 38399250, 38416365, 38445016, 38484295, 38593059, 38595656, 38649843, 38649864, 38745197, 38810587, 38810592, 38876171, 38876172, 38904806, 38937263, 38990883, 39092270, 39143220, 39189853, 39189855, 40485616, 40485617, 40492473, 40493384, 40497670, 40497857, 40499760, 40499761, 40502300, 40502325, 40507454, 40518477, 40518478, 40525595, 40525718, 40525719, 40536127;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0257-2027
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0257-2027. Report a data issue: hello@regdatalab.com.