Z-0260-2025 Class III Ongoing

Recalled by Trackx Technology Inc — Hillsborough, NC

FDA device recall Z-0260-2025 was initiated by Trackx Technology Inc on Sep 3, 2024 and is designated Class III. Reason for recall: The patient label stickers may have the wrong expiration date recorded. The contract manufacturer entered the manufacturing date (2024-04-01) as the expiration date. The actual expiration date of the… The recall status is ongoing. Affected quantity: 82.

Recall details

Product Type
Devices
Report Date
Nov 6, 2024
Initiation Date
Sep 3, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
82

Product description

Product Name: Insight Base Flex - 8, Ref 608-3TX; The device consists of a sterile Snap that includes reflective features that facilitate imaging of the (attached) instrument using instrument tracking technology when used with a compatible tracking system. Each TrackX¿ Snap has a unique set of 2 or more reflective markers. By having unique configurations of reflective markers, it is possible to identify which TrackX Snap is in the field of view of the camera(s). As such, it is also possible to have multiple TrackX Snaps in the field at once and to have different TrackX Snaps for different surgical instruments. The device is considered to be a sterile accessory.

Reason for recall

The patient label stickers may have the wrong expiration date recorded. The contract manufacturer entered the manufacturing date (2024-04-01) as the expiration date. The actual expiration date of the product is 2027-04-01. This is reflected on the carton label and sterile pouch label correctly.

Distribution pattern

NC, VA, FL, PA, NY

Code information

GTIN: 00857088008493, Lot: 2404002, Expiration Date: 04/01/2027

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0260-2025
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Trackx Technology Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Trackx Technology Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Trackx Technology Inc →

Source: openFDA record for Z-0260-2025. Report a data issue: hello@regdatalab.com.