Z-0262-2021 Class II Terminated

Recalled by ICU Medical, Inc. — San Clemente, CA

Recall Details

Product Type
Devices
Report Date
October 28, 2020
Initiation Date
July 6, 2020
Termination Date
May 11, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Total of all products (Listed #1 thru 101) = 304735 units

Product Description

8 IN (20CM) APPX 1.3ml, EXT SET w/3-Port NanoClave Manifold, Check Valve.1 unit per pouch, 50 pouches per case. UDI:(01)10887709085814(17)250101(30)50(10)4605676. CLAVE Neutron is a normally closed, bidirectional connector intended for use as an accessory to an Intravascular catheter placed in the vein or artery. The device may be used for the administration of blood and fluids to patients, including pediatrics and immunocompromised patients. .

Reason for Recall

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Distribution Pattern

Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.

Code Information

Lot Number:4605676.