Z-0264-2027 Class II Ongoing

Recalled by Restor3D Inc. — Durham, NC

FDA device recall Z-0264-2027 was initiated by Restor3D Inc. on Sep 4, 2026 and is designated Class II. Reason for recall: Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization. The recall status is ongoing. Affected quantity: 16.

Recall details

Product Type
Devices
Report Date
Oct 28, 2026
Initiation Date
Sep 4, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16

Product description

Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software Version: None Product Description: The restor3d Metatarsophalangeal (MTP) Implant is a sterile, single-use implant grade metal device, available in varied articulating surface coverages. The restor3d MTP implant is designed to replace the damaged articulating surface of the first metatarsal distal head. The device is intended for the treatment of patients with degenerative and post-traumatic arthritis in the first metatarsal joint in the presence of good bone stock and integrity of the first metatarsal head. This implant can also be used in the following clinical conditions: hallux valgus or hallux limitus, hallux rigidus, and an unstable or painful metatarsal/phalangeal (MTP) joint. The implant is a single use product intended to be press fit, with optional use of bone cement. The restor3d MTP Implant is comprised of a single laser powder bed fusion (LPBF) printed cobalt-chromium alloy (Co28Cr6Mo) part. The external surface of the implant that is intended to be in contact with the phalanx is mirror polished. Component: Not a component

Reason for recall

Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.

Distribution pattern

U.S. Nationwide distribution in the states of FL, IL, NC, PA, SC, and TX.

Code information

Lot Code: Model: 1103-161620 UDI: 00840097505244 Lot: A02979

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0264-2027
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.