FDA device recall Z-0265-2027 was initiated by Medtronic MiniMed, Inc. on Sep 2, 2026 and is designated Class II. Reason for recall: A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose… The recall status is ongoing. Affected quantity: 623.

Recall details

Product Type
Devices
Report Date
Oct 28, 2026
Initiation Date
Sep 2, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
623

Product description

Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601 (Android) Software Version: Software version 1.2.1 Product Description: The MiniMed Go app for iOS or MiniMed Go app for Android is compatible with the following components: " Simplera sensor " Instinct sensor " InPen smart insulin pen " Apple Watch [Optional] Continuous glucose monitoring (CGM) is a technology that allows you to continuously view glucose values. The app uses an inserted Simplera sensor CGM or Instinct sensor CGM with Bluetooth¿ wireless technology that converts small amounts of glucose from the interstitial fluid under the skin into an electronic signal. The sensors use that signal to provide sensor glucose (SG) values to the MiniMed Go app on a compatible mobile device. The app also provides alerts based on your SG levels and contains a user interface for entering data such as insulin, meals, exercise, blood glucose (BG) values, and uploading information to the CareLink" Personal website. The app is available to download from the Apple"* App Store"* or the Google Play"* Store. Component: No

Reason for recall

A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.

Distribution pattern

Worldwide - U.S. Nationwide distribution in the states of AL, CA, CT, FL, LA, MA, MD, NY, OH, PA, and TX. The countries of Austria, Belgium, Denmark, Finland, Germany, Iceland, Italy, Norway, Spain, Sweden, Switzerland, United Kingdom.

Code information

Lot Code: GTIN for MMT-8600, iOS: 00199150038978 GTIN for MMT-8601, Android: 00199150038985 UDI for MMT-8600, iOS: 00199150038978 UDI for MMT-8601, Android: 00199150038985 Software version 1.2.1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0265-2027
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.