Z-0266-2013 Class II Terminated
FDA device recall Z-0266-2013 was initiated by Stryker Howmedica Osteonics Corp. on Jul 3, 2012 and is designated Class II. Reason for recall: Three complaints have been filed where the Hoffman LRF Telescopic Struts have broken during load bearing application by patients having a body weight greater than 250 LBS. The recall status is terminated (terminated Apr 17, 2014). Affected quantity: 165 units.
Recall details
- Product Type
- Devices
- Report Date
- Nov 14, 2012
- Initiation Date
- Jul 3, 2012
- Termination Date
- Apr 17, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 165 units
Product description
Hoffman LRF Telescopic Strut Short Stryker Stryker Trauma AG CH-2545 Seizach Distributed in the USA by Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430. The use of external devices provides the surgeon with a means of bone fixation for the management of fracture and reconstructive surgery. These devices are intended only to assist healing and are not intended to replace normal bone structures. External fixation devices are intended to hold fractured bone surfaces in apposition to facilitate normal healing. While proven successful, the devices are manufactured from metal, plastic, and composite materials. No such fracture fixation device, subject to fatigue, can be expected to withstand activity levels in the same way as would a normal health bone. The external fixation system, therefore, will not be as strong, reliable or durable as a normal human bone.
Reason for recall
Three complaints have been filed where the Hoffman LRF Telescopic Struts have broken during load bearing application by patients having a body weight greater than 250 LBS.
Distribution pattern
Nationwide Distribution.
Code information
K113327 Catalog No: Telescopic Strut Extra Short Hoffman LRF Length: 4933-0-100 :100-125mm (Black) 4933-0-120 Telescopic Strut Short Hoffman LRF Length: 119-161 mm (Yellow) 4933-0-140 Telescopic Strut Medium Hoffman LRF Length 138-201 mm (Blue) 4933-0-180 Telescopic Strut Long Hoffman LRF Length: 177-277 mm (Red); Lot Code: 4933-0-100 Telescopic Strut Extra Short Hoffman LRF Length 100-125mm (Black) X28937; Z04197; Z09163; Z09597; Z12135; 4933-0-120 Telescopic Strut Short Hoffman LRF Length: 119-161 mm (Yellow) X28961; Z04199; Z09161; Z09769; Z12012; 4933-0-140 Telescopic Strut Medium Hoffman LRF Length 138-201 mm (Blue) X34352; Z04200; Z05068; Z09011; Z09598; Z10872; 4933-0-180 Telescopic Strut Long Hoffman LRF Length: 177-277 mm (Red); X28963 ;Z04429; Z09471; Z11686