Z-0267-2013 Class II Terminated

Recalled by Medacta Usa Inc — Camarillo, CA

Recall Details

Product Type
Devices
Report Date
November 14, 2012
Initiation Date
October 13, 2008
Termination Date
November 9, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product Description

Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.

Reason for Recall

Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.

Distribution Pattern

Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.

Code Information

Ref: 01.10.10.122 (Lot # 085119), Ref: 01.10.10.123 (Lot # 085120), Ref: 01.10.10.124 (Lot # 085244), Ref: 01.10.10.125 (Lot # 085243)