Z-0267-2013 Class II Terminated

Recalled by Medacta Usa Inc — Camarillo, CA

FDA device recall Z-0267-2013 was initiated by Medacta Usa Inc on Oct 13, 2008 and is designated Class II. Reason for recall: Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make… The recall status is terminated (terminated Nov 9, 2012). Affected quantity: 12 units.

Recall details

Product Type
Devices
Report Date
Nov 14, 2012
Initiation Date
Oct 13, 2008
Termination Date
Nov 9, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product description

Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.

Reason for recall

Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.

Distribution pattern

Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.

Code information

Ref: 01.10.10.122 (Lot # 085119), Ref: 01.10.10.123 (Lot # 085120), Ref: 01.10.10.124 (Lot # 085244), Ref: 01.10.10.125 (Lot # 085243)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0267-2013
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Medacta Usa Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Medacta Usa Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Medacta Usa Inc →

Source: openFDA record for Z-0267-2013. Report a data issue: hello@regdatalab.com.