Z-0272-2016 Class II Terminated

Recalled by Precision Medical, Inc. — Northampton, PA

FDA device recall Z-0272-2016 was initiated by Precision Medical, Inc. on Oct 8, 2015 and is designated Class II. Reason for recall: These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for… The recall status is terminated (terminated Oct 27, 2016). Affected quantity: 315.

Recall details

Product Type
Devices
Report Date
Nov 25, 2015
Initiation Date
Oct 8, 2015
Termination Date
Oct 27, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
315

Product description

Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.

Reason for recall

These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).

Distribution pattern

Distributed in the US to the states of PA, MA, MO, MT, ID, MI, MN, ND, KY, NY, NC, CA, AL, AZ, LA, OH, TX, GA, IN, and OK.

Code information

Lot numbers 072215 and 072315.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0272-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Precision Medical, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0272-2016. Report a data issue: hello@regdatalab.com.