Z-0278-2022 Class II Ongoing

Recalled by Philips North America Llc — Cambridge, MA

FDA device recall Z-0278-2022 was initiated by Philips North America Llc on May 1, 2021 and is designated Class II. Reason for recall: Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator… The recall status is ongoing. Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Dec 1, 2021
Initiation Date
May 1, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini GXL 16 Slice System Number: 882410 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.

Reason for recall

Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part

Distribution pattern

US (nationwide) distribution to states of: CA, ME, TX and OUS (Foreign) to countries of: Australia, Germany, Netherlands and Saudi Arabia.

Code information

Serial Number: 4165

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0278-2022
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips North America Llc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips North America Llc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips North America Llc →

Source: openFDA record for Z-0278-2022. Report a data issue: hello@regdatalab.com.