Z-0279-2013 Class II Terminated

Recalled by Terumo Cardiovascular Systems Corporation — Ann Arbor, MI

FDA device recall Z-0279-2013 was initiated by Terumo Cardiovascular Systems Corporation on Oct 16, 2012 and is designated Class II. Reason for recall: Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux… The recall status is terminated (terminated Jul 11, 2013). Affected quantity: 15 units.

Recall details

Product Type
Devices
Report Date
Nov 21, 2012
Initiation Date
Oct 16, 2012
Termination Date
Jul 11, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 units

Product description

Tested AC Control BD, 120V, X2 The Terumo HX2 Temperature Management System is a selfcontained, dual channel system for supplying temperature controlled water to heat exchangers used in an extracorporeal circuit.

Reason for recall

Based on a retrospective review of quality data, AC Control boards used in the manufacture of the HX2 Temperature Management System were identified as being manufactured using the incorrect flux material. Use of the incorrect flux may contribute to premature board failure. A search of TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred during set up

Distribution pattern

Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI

Code information

Catalog number: 817168 and serial numbers: 824POOO023, 831 P001130, 831 P001133, 831 P001136, 831 P001140, 831 POO1141 , 831 P001144, 831 P001145, 831 P001146, 831 P001149, 831 P001151, 831 P001159, 831P001161, 831P001164, and 836POOOO05.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0279-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Terumo Cardiovascular Systems Corporation? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Terumo Cardiovascular Systems Corporation: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Terumo Cardiovascular Systems… →

Source: openFDA record for Z-0279-2013. Report a data issue: hello@regdatalab.com.