Z-0279-2023 Class II Ongoing

Recalled by Southmedic, Inc. — Barrie, N/A

FDA device recall Z-0279-2023 was initiated by Southmedic, Inc. on Oct 27, 2022 and is designated Class II. Reason for recall: Upon review of its batch record, Southmedic identified that lot # W92761 was not sent out for sterilization, and was instead transferred directly to distribution where it was sold and distributed to… The recall status is ongoing. Affected quantity: 450 units.

Recall details

Product Type
Devices
Report Date
Nov 30, 2022
Initiation Date
Oct 27, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
450 units

Product description

Southmedic SAFETY CARTRIDGE WITH SOUTHMEDIC PLUS SURGICAL BLADE, REF 73-8020, STERILE EO

Reason for recall

Upon review of its batch record, Southmedic identified that lot # W92761 was not sent out for sterilization, and was instead transferred directly to distribution where it was sold and distributed to customers as non-sterile product labelled as sterile.

Distribution pattern

US Nationwide distribution in the state of IL.

Code information

GTIN 620974002628, Lot number: W92761

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0279-2023
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Southmedic, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0279-2023. Report a data issue: hello@regdatalab.com.