Z-0279-2023 Class II Ongoing

Recalled by Southmedic, Inc. — Barrie, N/A

Recall Details

Product Type
Devices
Report Date
November 30, 2022
Initiation Date
October 27, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
450 units

Product Description

Southmedic SAFETY CARTRIDGE WITH SOUTHMEDIC PLUS SURGICAL BLADE, REF 73-8020, STERILE EO

Reason for Recall

Upon review of its batch record, Southmedic identified that lot # W92761 was not sent out for sterilization, and was instead transferred directly to distribution where it was sold and distributed to customers as non-sterile product labelled as sterile.

Distribution Pattern

US Nationwide distribution in the state of IL.

Code Information

GTIN 620974002628, Lot number: W92761