Z-0280-2014 Class II Terminated

Recalled by Stryker Instruments Div. Of Stryker Corporation — Portage, MI

FDA device recall Z-0280-2014 was initiated by Stryker Instruments Div. Of Stryker Corporation on Nov 1, 2013 and is designated Class II. Reason for recall: On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522-000 Rev. H). On March 15, 2013, when the product was going to be placed in the pouch to… The recall status is terminated (terminated Feb 10, 2014). Affected quantity: 87 boxes (522 each).

Recall details

Product Type
Devices
Report Date
Nov 20, 2013
Initiation Date
Nov 1, 2013
Termination Date
Feb 10, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
87 boxes (522 each)

Product description

Acetabular Pressurizer; single use and six units per box. The Acetabular Pressurizer 0206-522-000 is an accessory to the Advanced Cement Mixer (ACM). The Acetabular Pressurizer Tiplet P/N 0206-522-012 is sealed in a two pouches, sterilized, and labeled as part number 0206-522-000. This Acetabular Pressurizer is an accessory used to pressurize the cement applied to the prosthesis to be implanted during the surgery. The Acetabular Pressurizer is an injection molded component made with resin material DOW HDPD 10462N which is a rigid plastic once it is molded.

Reason for recall

On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522-000 Rev. H). On March 15, 2013, when the product was going to be placed in the pouch to be sealed, the manufacturing operator noticed that some assemblies were cracked. As a result of the injection pressure of the cement gun, a cracked tiplet may release from the Hub during cement applica

Distribution pattern

Worldwide distribution: US (nationwide) including states of: TX, FL, IL, IA, ND, WI, VA, and NJ.; and countries of: CANADA, POLAND and SWITZERLAND.

Code information

Part number: 0206-522-000 and lots: 08183012, 08310012, 09237012, 10243012, 11109012, 11353012, and 12216012

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)