Z-0280-2022 Class II Ongoing
FDA device recall Z-0280-2022 was initiated by Philips North America Llc on May 1, 2021 and is designated Class II. Reason for recall: Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, b… The recall status is ongoing. Affected quantity: 3 units.
Recall Details
- Product Type
- Devices
- Report Date
- December 1, 2021
- Initiation Date
- May 1, 2021
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3 units
Product Description
Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini TF 16 System Number: 882470 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.
Reason for Recall
Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part
Distribution Pattern
US (nationwide) distribution to states of: CA, ME, TX and OUS (Foreign) to countries of: Australia, Germany, Netherlands and Saudi Arabia.
Code Information
Serial Number: 7084 7145 7584